Critical care : the official journal of the Critical Care Forum
-
Indirect calorimetry allows the determination of energy expenditure in critically ill patients by measuring oxygen consumption (VO2) and carbon dioxide production (VCO2). Recent studies have demonstrated variable performance of "breath-by-breath" instruments compared to mixing chamber technology. The aim of this study was to validate two modern devices (E-sCOVX and Quark RMR) against a reference method (Deltatrac II). ⋯ Both test devices overestimate VO2 and VCO2 compared to Deltatrac II. The observed limits of agreement are comparable to those commonly accepted in evaluations of circulatory monitoring, and significantly less than results from predictive equations. We hypothesize that the discrepancy between methods is due to patient/ventilator-related factors that affect the synchronization of gas and spirometry waveforms.
-
The 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation recommend Basic Life Support (BLS) and Advanced Life Support (ALS) rules for termination of resuscitation (TOR). However, it is unclear whether the TOR rules are valid for out-of-hospital cardiac arrests (OHCAs) of both cardiac and non-cardiac etiologies. In this study, we validated the TOR rules for OHCA resulting from both etiologies. ⋯ Both TOR rules have high specificity and PPV in patients with OHCA from cardiac etiologies. For patients with OHCA from non-cardiac etiologies, the rules had a high PPV, but relatively low specificity. Therefore, TOR rules are useful in patients with OHCA from cardiac etiologies, but should be applied with caution to patients with OHCA from non-cardiac etiologies.