Journal of clinical monitoring and computing
-
J Clin Monit Comput · Dec 2020
Does it make difference to measure diaphragm function with M mode (MM) or B mode (BM)?
Diaphragm dysfunction occurs in mechanically ventilated subjects. Recent literature suggests that diaphragm thickening fraction (DTF) measured by ultrasound can be useful to predict weaning outcome. However, there is no standardized approach in the measurement of diaphragm thickness (DT) and limited data exists comparing different measurement techniques of diaphragm thickness (M mode-MM or B mode-BM). ⋯ BM and MM tidal diaphragm measurements during the inspiratory (0.3 ± 0.08 and 0.31 ± 0.08 cm; P = 0.022), expiratory (0.24 ± 0.07 and 0.24 ± 0.07 cm; P = 0.315) phases and tidal DTF were (27 ± 16 and 31 ± 14%, P = 0.089) respectively. Results of our study suggests that except tidal inspiratory diaphragm thickness, all thickness and excursion measurements with MM and BM are very compatible with each other. Further studies are necessarry to confirm our results and to standardize the measurements of diaphragm.