Respiratory care
-
Positive expiratory pressure (PEP) therapy imposes expiratory flow resistance to increase airway diameter and enhance mucus clearance. PEP is achieved several ways. Oscillatory PEP devices (OPEP) generate repeated occlusions that are known to reduce mucus viscosity. There are many marketed devices, but comparative performance is mostly unreported. The purpose of this study was to evaluate performance characteristics of many PEP/OPEP devices. For OPEP devices, we defined an optimal performance metric by creating an oscillation index that combines the OPEP performance characteristics. ⋯ PEP devices behaved similarly and as expected, with increased pressure with increased flow (flow resistors) or flow independence (threshold resistors). There was much greater variation in the performance of the OPEP devices. A higher oscillation index indicates better mechanical performance characteristics. Many devices have similar characteristics. However, the devices with the highest oscillation index have the highest flow amplitude and frequency, which may indicate better clinical performance.
-
Editorial Comment
Ventilator Shortages and Solutions, Real and Imagined.