Respiratory care
-
The aim of this study was to investigate the effects of different delivery circuit configurations, nebulizer positions, CPAP levels, and gas flow on the amount of aerosol bronchodilator delivered during simulated spontaneous breathing in an in vitro model. ⋯ Based on our in vitro study, without CPAP, a T-piece with a cap at one extremity maximizes albuterol delivery. During high-flow CPAP, the nebulizer should always be placed proximal to the patient, after the T-piece, using the highest CPAP clinically indicated.
-
Bubble CPAP, used for spontaneously breathing infants to avoid intubation or postextubation support, can be delivered with different interface types. This study compared the effect that interfaces had on CPAP delivery. We hypothesized that there would be no difference between set and measured levels between interface types. ⋯ Set flow had a profound effect on measured CPAP. The concomitant drop in measured pressure with high and low flows could be attributed to increased resistance to spontaneous breathing or insufficient flow to meet inspiratory demand. Clinicians should be aware of the effect that the interface and flow have on CPAP delivery.
-
High-flow nasal cannula (HFNC) therapy provides better humidification than conventional oxygen therapy. To allay loss of vapor as condensation, a servo-controlled heating wire is incorporated in the inspiratory tube, but condensation is not completely avoidable. We investigated factors that might affect condensation: thermal characteristics of the inspiratory tube, HFNC flow, and ambient temperature. ⋯ Ambient temperature statistically significantly influenced the amount of condensation in the tubes.