Rocuronium produces faster neuromuscular blockade compared with other neuromuscular blocking drugs. It produces comparable intubating conditions to that of succinylcholine, but does not have the short intubation time of the latter. Hence, it may not be preferable for rapid sequence intubation, but rocuronium with priming may produce comparable intubating time and conditions to that of succinylcholine. Rocuronium with priming may be an alternative to succinylcholine in rapid sequence intubation in conditions where succinylcholine is contraindicated. The present study was conducted to compare the intubating conditions and intubation time of rocuronium with and without priming. ⋯ Priming with rocuronium provides excellent intubating conditions in less than 60 s with no adverse effects.
-
Optimal depth of endotracheal tube (ET) placement has been a serious concern because of the complications associated with its malposition. ⋯ Fixing the tube at recommended 23 cm in males and 21 cm in females will lead to carinal stimulation or endobronchial placement in many Indian patients. The lip to carina distance best correlates with patient's height. Positioning the ET tip 4 cm above carina as recommended will result in placement of tube cuff inside cricoid ring with currently available tubes. Optimal depth of ET placement can be estimated by the formula "(Height in cm/7)-2.5."