Respiratory mucosal and lung structures and functions may be severely impaired in mechanically ventilated patients when delivered gases are not adequately conditioned. Although under- and over-humidification of respiratory gases have not been defined clearly, a safe range of temperature and humidity may be suggested. During mechanical ventilation, gas entering the trachea should reach at least physiologic conditions (32 degrees C-34 degrees C and 100%relative humidity) to keep the ISB at its normal location. ⋯ Full saturation of inspiratory gases is likely when water condensation is observed in the flex tube [91,92]. Nevertheless, no clinical parameter is accurate enough to detect all the effects of inadequate conditioning [45]. When mechanical ventilation is extended beyond several days, adequate conditioning of respiratory gases becomes increasingly crucial to prevent retention of secretions and to maximize mucociliary function; a requirement that respiratory gases reach at least physiologic conditions is appropriate.