Journal of clinical monitoring
-
The most efficient site for monitoring heart and lung sounds by esophageal stethoscope is not the warmest segment of the esophagus. This study investigated the ability of passive warming of airway gases to increase the accuracy of temperatures measured at this site (i.e., to decrease their difference from core temperature). In 15 adult patients undergoing general anesthesia and endotracheal intubation, esophageal temperatures were measured before and after use of a heat and moisture exchanger (an artificial nose) that passively warmed inspired gases. ⋯ After passive warming of inspired gases, esophageal temperatures increased significantly (mean increase +/- SD, 0.5 +/- 0.2 degrees C; P less than or equal to 0.001) but inconsistently (range, 0.1 to 1.2 degrees C). However, the mean difference between esophageal and nasopharyngeal temperatures was still significant (0.5 +/- 0.3 degrees C; P less than 0.001). Discrepancies between esophageal and core temperatures persist when a currently available esophageal stethoscope with adjacent auscultation chamber and temperature probe is used, despite passive warming of airway gases.