We evaluated the short-term variability of PaO2, PaCO2, pulse oximeter saturation (SpO2), and end-tidal PCO2 (PETCO2) in mechanically ventilated trauma patients. All patients were stable and undisturbed during the evaluation periods. Blood gases were obtained from an arterial catheter 4 times at 20-minute intervals. ⋯ The variability of PETCO2 was similar to the variability of PaCO2. However, the variability of PaO2 was considerably greater than that of SpO2, which was probably related to the shape of the oxyhemoglobin dissociation curve and the relatively high saturations of the patients in this study. Variability of blood gases, SpO2, and PETCO2 should be considered when these values are clinically interpreted.
-
We developed a noninvasive computer-based system for estimating continuous cardiac output by a modified pulse contour method using a finger pressure waveform. The method requires no individual patient calibration or baseline cardiac output. First, we calibrated the system in a "learn" group of 20 patients. ⋯ The program demonstrated that information about cardiac output can be obtained by using the Finapres device (Ohmeda, Boulder, CO). The cardiac output values obtained by this continuous noninvasive technique were within +/- 20% of the simultaneous thermodilution values 87% of the time. This was true over the narrow range of cardiac outputs (2.9 to 6.4 L/min) and wide range of heart rates (45 to 140 beats/min).