Fixed performance oxygen masks operate by supplying mixtures of oxygen and air at rates exceeding the inspiratory flow rate of the patient. In this study the oxygen concentration delivered by three fixed performance oxygen masks was determined non-invasively at various inspiratory flow rates. ⋯ The change from fixed to variable performance depended on the relation between inspiratory flow rate and the total gas flow delivered by the mask and was independent of the volume of the mask. Hence the use of low volume masks and high oxygen flow rates should produce more consistent results than high volume masks and lower flow rates.
-
Br Med J (Clin Res Ed) · Mar 1984
Randomized Controlled Trial Clinical TrialEndotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.
Large volume, low pressure endotracheal tube cuffs are claimed to have less deleterious effect on tracheal mucosa than high pressure, low volume cuffs. Low pressure cuffs, however, may easily be overinflated to yield pressures that will exceed capillary perfusion pressure. Various large volume cuffed endotracheal tubes were studied, including Portex Profile, Searle Sensiv, Mallinkrodt Hi-Lo, and Lanz. ⋯ It was found that these cuffs when overpressurised impaired mucosal blood flow. This impairment of tracheal mucosal blood flow is an important factor in tracheal morbidity associated with intubation. Hence it is recommended that a cuff inflation pressure of 30 cm H2O (22 mm Hg) should not be exceeded.