Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
-
Non-transvenous implantable cardioverter-defibrillators (ICDs) as used in small patients and in patients with congenital heart disease (CHD) have not been compared with transvenous systems with respect to safety and efficacy yet. Aim of the present study was to describe the prevalence of and to identify contributing factors for appropriate and inappropriate ICD discharges in patients with non-transvenous and transvenous ICD. ⋯ Individuals with EC-ICD systems were particularly prone to experience appropriate shocks. As rapidly conducted AT was the most common reason for inappropriate ICD shocks, rigorous treatment of these arrhythmias and proper ICD programming are mandatory. Though lead failure was of concern in EC-ICD patients, EC-ICD systems were not inferior with respect to inappropriate shocks.
-
We have recently shown that a contact force (CF)-guided ablation protocol respecting region-specific criteria of lesion contiguity and lesion depth ('CLOSE' protocol) is associated with high incidence of acute durable pulmonary vein (PV) isolation (PVI) and a high single-procedure arrhythmia-free survival at 1 year. In the present study, we compared efficiency, safety, and efficacy of 'CLOSE'-guided PVI to conventional CF-guided PVI (CONV-CF). ⋯ 'CLOSE'-guided PVI improves procedural and 1 year outcome in CF-guided PVI while shortening procedure time. Improvement cannot be explained by differences in CF variability and is most likely due to the strict application of criteria for contiguity and ablation index. A randomized controlled trial is needed to exclude the possible contribution of a learning curve.
-
Magnetic resonance imaging (MRI) is the gold standard for defining myocardial substrate in 3D and can be used to guide ventricular tachycardia ablation. We describe the feasibility of using a prototype magnetic resonance-guided electrophysiology (MR-EP) system in a pre-clinical model to perform real-time MRI-guided epicardial mapping, ablation, and lesion imaging with active catheter tracking. ⋯ The MR-EP system was able to successfully acquire epicardial voltage and activation maps in swine, deliver, and visualize ablation lesions, demonstrating feasibility for intraprocedural guidance and real-time assessment of ablation injury.
-
Review
The athlete's heart is a proarrhythmic heart, and what that means for clinical decision making.
Recurring questions when dealing with arrhythmias in athletes are about the cause of the arrhythmia and, more importantly, about the eligibility of the athlete to continue sports activities. In essence, the relation between sports and arrhythmias can be understood along three lines: sports as arrhythmia trigger on top of an underlying problem, sports as arrhythmic substrate promotor, or sports as substrate inducer. Often, there is no sharp divider line between these entities. ⋯ This review describes the changes in the athlete's heart that are related to arrhythmic expression and focuses on what this concept means for clinical decision making. The concept of the athlete's heart as a proarrhythmic heart creates a framework for evaluation and counselling of athletes, yet also highlights the difficulty in predicting the magnitude of associated risk. The management uncertainties are discussed for specific conditions like extreme bradycardic remodelling, atrioventricular nodal reentrant tachycardia, atrial fibrillation and flutter, and ventricular arrhythmias.
-
Comparative Study
Propensity score matched comparison of subcutaneous and transvenous implantable cardioverter-defibrillator therapy in the SIMPLE and EFFORTLESS studies.
Comparison of outcomes between subcutaneous and transvenous implantable cardioverter-defibrillator (S-ICD and TV-ICD) therapy is hampered by varying patient characteristics and complication definitions. The aim of this analysis is to compare clinical outcomes of S-ICD and TV-ICD therapy in a matched cohort. ⋯ The earliest experience of the S-ICD demonstrates similar outcomes as contemporary TV-ICD therapy in a matched comparison with predominately single-chamber devices at 3 years follow-up.