Researchers induced awake paralysis in 10 volunteers using separately both suxamethonium and rocuronium. Both the BIS A2000 (2003) and BIS Vista monitor (2013) were tested.
BIS decreased immediately after paralysis and did not fully recover until muscle recovery. BIS values decreased to as low as 44, despite the subject being awake.
In more than half of the 20 trials the BIS value decreased to below 60 at some point. In one case this lasted for almost 4 minutes, representing 76% of the total paralysis time for that subject.
Interesting editorial accompanying Dr Peter Schuller's excellent study of BIS values in awake, paralysed volunteers.
The editors make a very interesting point critiquing the probabilistic, database-based approach to processed-EEG awareness monitors like BIS: (emphasis added)
"This database-driven approach may have limitations, in particular for the detection of intraoperative wakefulness: it is very unlikely that data from an awake and paralyzed subject are included in this database. Therefore, the resulting anaesthesia index has not been trained with a dataset that contains this clinical situation..."