• Neuromodulation · Jan 2013

    Impedance variations over time for a closed-loop neurostimulation device: early experience with chronically implanted electrodes.

    • Chengyuan Wu, James J Evans, Christopher Skidmore, Michael R Sperling, and Ashwini D Sharan.
    • Department of Neurological Surgery, Thomas Jefferson University, Philadelphia, PA 19107, USA. chengyuan.wu@jeffersonhospital.org
    • Neuromodulation. 2013 Jan 1;16(1):46-50; discussion 50.

    ObjectiveResponsive neurostimulation (RNS®) is an investigational treatment modality for intractable focal epilepsy. We analyzed impedance values over time to gain a better understanding of RNS device stability, potential differences between depth and strip electrodes, and general implications of long-term electrode implantation.Materials And MethodsImpedance measurements were retrospectively reviewed from seven patients over a 3-year period. Statistical analysis was performed to investigate trends in impedance values over time and to determine if any differences existed between the means and variances of impedance values for the two different electrodes.ResultsIntraparenchymal depth electrodes demonstrated lower mean impedances and had less variation over time when compared to subdural strip electrodes. There was no significant change in mean impedance over time for depth electrodes, while that of subdural electrodes increased over time to peak at 16-20 months and returned to baseline by 2 years.ConclusionThe RNS device provides unique long-term data for both depth and subdural electrodes in human subjects. Although changes in impedance were observed with respect to time after implantation and electrode type, these variations were of a clinically insignificant magnitude. As a constant-current system, the device is ultimately able to administer stable therapeutic doses in the setting such minor variability. While further animal research and post mortem investigations in human subjects are necessary to broaden our understanding of the mechanisms occurring at the electrode-neural interface and the biocompatibility of chronically implanted electrodes, the relative stability of impedances seen in this cohort of chronic implants brings promise to future implementations of chronic intracranial neural implants.© 2012 International Neuromodulation Society.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.