• Resuscitation · Aug 2020

    Chest compressions induce errors in end-tidal carbon dioxide measurement.

    • Mikel Leturiondo, Sofía Ruiz de Gauna, José Julio Gutiérrez, Daniel Alonso, Carlos Corcuera, Juan Francisco Urtusagasti, Digna María González-Otero, James Knox Russell, Mohamud Ramzan Daya, and Jesus María Ruiz.
    • University of the Basque Country, UPV/EHU, Bilbao, Bizkaia, Spain. Electronic address: mikel.leturiondo@ehu.eus.
    • Resuscitation. 2020 Aug 1; 153: 195-201.

    BackgroundReal-time measurement of end-tidal carbon dioxide (ETCO2) is used as a non-invasive estimate of cardiac output and perfusion during cardiopulmonary resuscitation (CPR). However, capnograms are often distorted by chest compressions (CCs) and this may affect ETCO2 measurement. The aim of the study was to quantify the effect of CC-artefact on the accuracy of ETCO2 measurements obtained during out-of-hospital manual CPR.MethodsWe retrospectively analysed monitor-defibrillator recordings collected by two advanced life support agencies during out-of-hospital cardiac arrest. These two agencies, represented as A and B used different side-stream capnometers and monitor-defibrillators. One-minute capnogram segments were reviewed. Each ventilation within each segment was identified using the transthoracic impedance signal and the capnogram. ETCO2 values per ventilation were manually annotated and compared to the corresponding capnometry values stored in the monitor-defibrillator. Ventilations were classified as distorted or non-distorted by CC-artefact.ResultsA total of 407 1-min capnogram segments from 65 patients were analysed. Overall, 4095 ventilations were annotated, 2170 (32.4% distorted) and 1925 (31.8% distorted) for agency A and B, respectively. Median (IQR) unsigned error in ETCO2 measurement increased from 1.5 (0.6-3.1)% for non-distorted to 5.5 (1.8-14.1)% for distorted ventilations; from 0.7 (0.3-1.2)% to 3.7 (1.0-9.9)% in agency A and from 2.3 (1.2-3.9)% to 8.3 (3.9-19.5)% in agency B (p < 0.001). Errors were higher than 10 mmHg in 9% and higher than 15 mmHg in 5% of the distorted ventilations.ConclusionCC-artefact causes ETCO2 measurement errors in the two studied devices. This suggests that capnometer algorithms may need to be adapted to reliably perform in the presence of CC-artefact during CPR.Copyright © 2020 Elsevier B.V. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…