• Respiratory care · Jan 2021

    Measured continuous positive airway pressure in a non-invasive pediatric airway and lung model.

    • Neil D Fernandes, Esther Chung, Michael D Salt, Beverly Ejiofor, Ryan W Carroll, and Robert M Kacmarek.
    • Massachusetts General Hospital, Boston, Massachusetts. ndfernandes@mgh.harvard.edu.
    • Respir Care. 2021 Jan 1; 66 (1): 879487-94.

    BackgroundBronchiolitis is the most common cause of admission in children < 2 y of age in the United States. The standard of care involves supportive measures, including noninvasive interventions such as CPAP. CPAP is traditionally delivered through a full face mask; however, pediatric ICUs have been exploring the use of the RAM cannula by Neotech as a mode of CPAP delivery. The level of CPAP delivered via the RAM cannula is uncertain. We performed an in vitro study to determine the level of CPAP delivered via the RAM cannula utilizing a pediatric lung model.MethodsModels of 7 sizes of pediatric upper airways, produced with a 3-dimensional printer, were connected to a breathing simulator. We applied each size of RAM cannula to weight-appropriate airway and lung compliance parameters, delivering pressures of 5, 7, and 10 cm H2O using a ventilator in the CPAP mode. Leaks of 0%, 20%, 40%, and 60% were generated to emulate a complete seal, a poor fit, and open-mouth breathing. The outcome measure was the difference in CPAP, referred to as "%leak effect," measured by the lung simulator relative to the CPAP set on the ventilator.ResultsWe found that set CPAP of 5-10 cm H2O generated measured CPAP ranging from 2.6 to 9.7 cm H2O. For the set CPAP levels of 5, 7, and 10 cm H2O, the mean %leak effect values of measured CPAP from the set CPAP were -25%, -26%, and -25.7%, respectively. For each specific cannula-airway combination, increasing the set pressure and decreasing the air leak resulted in higher levels of CPAP delivered.ConclusionsThe RAM cannula delivered varying amounts of CPAP, with a percent loss of approximately -25% depending on the level of leak in the system. With minimal leak, it is conceivable that the RAM cannula can be used to deliver clinically meaningful CPAP.Copyright © 2021 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…