    A Comprehensive Bench Assessment of Automatic Tube Compensation in ICU Ventilators for Better Clinical Management.

    • Louis-Marie Galerneau, Nicolas Terzi, Emanuele Turbil, Zakaria Riad, Carole Schwebel, Martin Cour, Laurent Argaud, Claude Guérin, and Bruno Louis.
    • Médecine Intensive Réanimation, Université de Grenoble-Alpes, Grenoble, France.
    • Respir Care. 2020 Dec 1; 65 (12): 1864-1873.

    BackgroundAutomatic tube compensation (ATC) unloads endotracheal tube (ETT) resistance. We conducted a bench assessment of ATC functionality in ICU ventilators to improve clinical management.MethodsThis study had 2 phases. First, we performed an international survey on the use of ATC in clinical practice, hypothesizing a rate of ATC use of 25%. Second, we tested 7 modern ICU ventilators in a lung model mimicking a normal subject (Normal), a subject with ARDS, and a subject with COPD. Inspiratory effort consisted of esophageal pressure over 30 consecutive breaths obtained in a real patient under weaning. A brand new 8-mm inner diameter ETT was attached to the lung model, and ATC was set at 100% compensation for the ETT. The 30 breaths were first run with ATC off and no ETT (ie, reference period), and then with ATC on and ETT (ie, active period). The primary end point was the difference in tidal volume (VT) between reference and active periods. We hypothesized that the VT difference should be equal to 0 in an ideally functioning ATC. VT difference was compared across ventilators and respiratory mechanics conditions using a linear mixed-effects model.ResultsThe clinical use of ATC was 64% according to 644 individuals who responded to the international survey. The VT difference varied significantly across ventilators in all respiratory mechanics configurations. The divergence between VT difference and 0 was small but significant: the extreme median (interquartile range) values were -0.013 L (-0.019 to -0.002) in the COPD model and 0.056 L (0.051-0.06) in the Normal model. VT difference for all ventilators was 0.015 L (95% CI 0.013-0.018) in the ARDS model, which was significantly different from 0.021 L (95% CI 0.018-0.024) in the Normal model (P < .001) and 0.010 L (0.007-0.012) in the COPD model (P = .003).ConclusionsATC is used more frequently in clinical practice than expected. In addition, VT delivery by ATC differed slightly though significantly between ventilators.Copyright © 2020 by Daedalus Enterprises.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.