• Neuromodulation · Apr 2006

    Factors affecting impedance of percutaneous leads in spinal cord stimulation.

    • Kenneth Alò, Clayton Varga, Elliot Krames, Joshua Prager, Jan Holsheimer, Ljubomir Manola, and Kerry Bradley.
    • Pain and Health Management Center, Houston, Tex; Pasadena Rehabilitation Institute, Pasadena, Calif; Pacific Pain Treatment Center, San Francisco, Calif; California Pain Medicine Center, Los Angeles, Calif, USA; University of Twente, Enschede, The Netherlands; Advanced Bionics Corp., Valencia, Calif, USA.
    • Neuromodulation. 2006 Apr 1;9(2):128-35.

    AbstractObjectives.  Although the load impedance of a pulse generator has a significant effect on battery life, the electrical impedance of contact arrays in spinal cord stimulation (SCS) has not been extensively studied. We sought to characterize the typical impedance values measured from common quadripolar percutaneous SCS contact arrays. Methods.  In 36 patients undergoing percutaneous trial stimulation for various chronic pain conditions, bipolar impedance between adjacent contacts of 64 leads with 9 mm center-to-center spacing was measured in two different vertebral level regions, cervical (C3-C7) and lower-thoracic (T7-T12). Multiple linear regression was applied to analyze the contribution of six variables to the biological tissue portion of the impedance (excluding the resistance of the lead wires). Results.  The median impedance in the cervical region (351 ± 90 Ω) was significantly lower (36%, p < 0.001) than in the lower-thoracic region (547 ± 151 Ω). In addition, time since implant had a weaker but still significant effect on tissue impedance. Conclusions.  Results from finite-difference mathematical modeling of SCS suggest that the difference in tissue impedance related to vertebral level may be due to the dorsoventral position of the lead in the epidural space. The presence of a larger space between the triangularly shaped dorsal part of the vertebral arch and the round shape of the dural sac in the lower-thoracic region increases the likelihood that the stimulating lead will not make dural contact, and thus "see" an increased impedance from the surrounding epidural fat. This implies that the energy requirements for stimulation in the thoracic region will be higher than in the cervical region, at least during the acute phase of implant.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…