• Acta Neurochir. Suppl. · Jan 2012

    Telemetric ICP measurement with the first CE-approved device: data from animal experiments and initial clinical experiences.

    • Michael Kiefer, Sebastian Antes, Steffen Leonhardt, Melanie Schmitt, Berk Orakcioglu, Oliver W Sakowitz, and Regina Eymann.
    • Department of Neurosurgery, Saarland University, Homburg-Saar, Germany. michael.kiefer@uks.eu
    • Acta Neurochir. Suppl. 2012 Jan 1; 114: 111-6.

    AbstractThe objective was to evaluate the qualification of the new telemetric intracranial pressure (ICP) measurement (t-ICP) device Raumedic(®) NEUROVENT P-Tel and S-Tel. The proof of concept was examined in a pilot animal study measuring intraperitoneal pressure with a telemetric and a conventional ICP measurement probe at five rates for 1 h each. Moderate external pressure load allowed measuring values between 0 and 40 mmHg. To estimate long-term performance 18 t-ICP devices were implanted subdurally or intraparenchymally into minipigs. Reference measurements were performed regularly using conventional ICP probes. From the short-term as well as from the long-term perspective t-ICP proved to have excellent dynamic ICP signal components perception (e.g. pulse amplitude). Some zero drift of static ICP was found, ranging between 5 and 8 mmHg. While all telemetric, intraparenchymal probes kept their functionality throughout the follow-up, 33% of the subdurals failed for reasons detailed in another paper. Raumedic's NEUROVENT(®) P-Tel/S-Tel proved to provide reliable data over periods of up to 18 months. Minor zero drift can be well tolerated as the dynamic ICP signal is measured with excellent stability. Clinicians should focus more on such ICP dynamic signal information than on static ICP when using the device over longer follow-up periods.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.