ObjectivesThree-dimensional (3D) transesophageal echocardiography (TEE) has been shown to be more accurate than 2D TEE for the evaluation of the left ventricular outflow tract area. The aim of the present study was to compare the agreement of 3D echocardiography-derived cardiac output (CO) with thermodilution-derived CO (TDCO) before and after cardiopulmonary bypass (CPB).DesignThis was a prospective observational study of patients who underwent cardiac surgery between 2016 and 2018.SettingWeill Cornell Medicine, a single large academic medical center.ParticipantsThe study comprised 78 patients undergoing elective cardiac surgery.InterventionsCPB, TEE, pulmonary artery catheter, and elective cardiac surgery.Measurements And Main ResultsTwo-dimensional CO, 3D CO-diameter, and 3D CO-area values pre-CPB were strongly correlated with one another both pre-CPB and post-CPB. The 3D CO-diameter and the 3D CO-area were mildly correlated, with TDCO measurements pre-CPB (r = 0.46 and 0.39, respectively) and post-CBP (r = 0.43 and 0.47, respectively). Pre-CPB 3D CO-diameter had the most agreement with TDCO in terms of bias (-0.13 L/min); however, the limits of agreement (LOA) were wide (-2.2- to- 2.45 L/min). Post-CPB, 3D CO-diameter had the most agreement with TDCO in terms of bias (0.41) but with wide LOA (-3.29 to 2.47). All pre-CPB echocardiography-derived CO (2D CO, 3D CO-diameter, 3D CO-area) had more agreement with TDCO than did post-CPB measurements.ConclusionsThree-dimensional CO measurements were only modestly correlated with pulmonary artery catheter-derived CO pre-bypass and post-bypass. Despite low bias, the wide LOA from 2D CO, 3D CO-diameter, and 3D-area compared with TDCO suggested that the 2 methods are not interchangeable.Copyright © 2020 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.