Validation of noninvasive continuous arterial pressure measurement by ClearSight System™ during induction of anesthesia for cardiovascular surgery.
- Tadashi Tanioku, Akari Yoshida, Yuichi Aratani, Keisuke Fujii, and Tomoyuki Kawamata.
- Department of Anesthesiology, Wakayama Medical University School of Medicine, Kimiidera 811-1, Wakayama, 641-8509, Japan. ttanioku@wakayama-med.ac.jp.
- BMC Anesthesiol. 2020 Jul 20; 20 (1): 176.
BackgroundSince blood pressure tends to be unstable during induction of anesthesia in patients undergoing cardiovascular surgery, an artery catheter is often inserted before induction to continuously monitor arterial pressure during induction of anesthesia. ClearSight System™ enables noninvasive continuous measurement of beat-to-beat arterial pressure via a single finger cuff without pain using photoplethysmographic technology. If ClearSight System™ can replace intra-arterial pressure measurement, blood pressure could be easily and noninvasively assessed. However, the validity of ClearSight System™ during induction of anesthesia in patients undergoing cardiovascular surgery has not been evaluated. The aim of this study was to compare blood pressure measured by ClearSight System™ with intra-arterial pressure during induction of anesthesia for cardiovascular surgery.MethodsThis study was registered retrospectively. Data during induction of anesthesia for elective cardiovascular surgery were obtained for patients in whom noninvasive arterial pressure was measured by ClearSight System™ (APcs) and invasive radial arterial pressure (APrad) was measured simultaneously. According to the widely used criteria formulated by international standards from the Association for the Advancement of Medical Instrumentation, the acceptable bias and precision for arterial pressure measurements were fixed at < 5 mmHg and 8 mmHg, respectively.ResultsData for 18 patients were analyzed. For 3068 analyzed paired measurements, values of APcs vs APrad bias (precision) were 13.2 (17.5), - 9.1 (7.3) and - 3.9 (7.8) mmHg for systolic, diastolic, and mean arterial pressures, respectively.ConclusionsMean arterial pressure measured by ClearSight System™ could be considered as an alternative for mean radial arterial pressure during induction of anesthesia for elective cardiovascular surgery.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.