Comparative Study
Comparison of Vibrating Mesh, Jet, and Breath-Enhanced Nebulizers During Mechanical Ventilation.
- Sunya Ashraf, Michael McPeck, Ann D Cuccia, and Gerald C Smaldone.
- Division of Pulmonary, Critical Care and Sleep Medicine, Department of Medicine, Stony Brook University Medical Center, Stony Brook, New York. sunya.ashraf@stonybrookmedicine.edu.
- Respir Care. 2020 Oct 1; 65 (10): 1419-1426.
BackgroundThis study compared 3 nebulizer technologies for inter- and intradevice reproducibility, humidification, and fill volume sensitivity during mechanical ventilation: a breath-enhanced jet nebulizer, a vibrating mesh nebulizer, and a jet nebulizer. The breath-enhanced jet nebulizer featured a new design located on the wet side of the humidifier to reduce aerosol loss and potential humidifier contamination. The vibrating mesh nebulizer and the jet nebulizer were placed on the dry side.MethodsAerosol delivery was measured using multiple ventilator settings (inspiratory time = 0.45-1.01 s). Using radiolabeled saline and a gamma camera, bench studies were performed using a ventilator to test 4 breathing patterns. Four scenarios were assessed during testing: 3 mL and 6 mL fill volumes with and without heated wire humidification. Measurements included inhaled mass (as a percentage of the nebulizer charge), nebulizer residual, mass balance, and aerosol particle size distribution. Statistics were determined using Mann-Whitney and linear regression.ResultsThe inhaled mass for the breath-enhanced jet nebulizer was 10.5-29.2% and was affected by fill volume (P = .004) but not by humidity. The inhaled mass for the vibrating mesh nebulizer was 0.9-33% and was unaffected by fill volume and humidity. The inhaled mass for the jet nebulizer was 2.5-25.9% and was affected by both fill volume (P = .009) and humidity (3 mL, P = .002). The inhaled mass for the vibrating mesh nebulizer was more variable due to random failures to achieve complete nebulization, and inhaled mass correlated closely with residual mass: IM% = -0.233(Residual%) + 24.3, r2 = 0.67, P < .001. For all devices, large particles were lost in the ventilator tubing; large particles were also lost in the humidifier for the vibrating mesh nebulizer (17% nebulizer charge), resulting in similar particle distributions (mass median aerodynamic diameter 1.33-1.95 μm) for all devices.ConclusionsNebulization with the breath-enhanced jet nebulizer was less sensitive to humidification than the jet nebulizer. Delivery via the vibrating mesh nebulizer was not predictable, with random failure to empty (55% experimental runs). All devices delivered similar particle distributions. Wet-side aerosol delivery avoids humidifier contamination, and breath-enhanced technology can ensure better control of drug delivery.Copyright © 2020 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.