• BJU international · Apr 2016

    Urethral atrophy after implantation of an artificial urinary sphincter: fact or fiction?

    • Simon Bugeja, Stella L Ivaz, Anastasia Frost, Daniela E Andrich, and Anthony R Mundy.
    • Reconstructive Urology Unit, University College London Hospitals, London, UK.
    • BJU Int. 2016 Apr 1; 117 (4): 669-76.

    ObjectivesTo investigate the concept of 'urethral atrophy', which is often cited as a cause of recurrent incontinence after initially successful implantation of an artificial urinary sphincter (AUS); and to investigate the specific cause of the malfunction of the AUS in these patients and address their management.Patients And MethodsBetween January 2006 and May 2013, 50 consecutive patients (mean age 54.3 years) with recurrent incontinence had their AUS explored for malfunction and replaced with a new device with components of exactly the same size, unless there was a particular reason to use something different. Average time to replacement of the device was 10.1 years. The mean follow-up after replacement of the device was 24.7 months. All patients without an obvious cause for their recurrent incontinence had preoperative urodynamic evaluation, including measurement of the Valsalva leak point pressure (VLPP) and the retrograde cuff occlusion pressure (RCOP). After explantation of the AUS in patients without any apparent abnormality of the device at the time of replacement, the pressure generated by the explanted pressure-regulating balloon (PRB) was measured manometrically, when this was possible. In a select group of six consecutive patients of this type, the fibrous capsule surrounding the old cuff was incised then excised to expose and evaluate the underlying corpus spongiosum.ResultsIn 31 of the 50 patients (62%) undergoing exploration, a specific cause for the malfunction of their AUS was defined. In the other 19 patients (38%) no cause was found, either preoperatively or at the time of exploration, other than a low VLPP and RCOP. A typical 'waisted' or 'hour-glass' appearance of the underlying corpus spongiosum was demonstrable, to some degree, on explanting the cuff in all cases. In the six patients in whom the restrictive sheath surrounding the cuff was excised, the urethral circumference immediately returned to normal after the compressive effect of the sheath was released. Manometry of the explanted PRBs, when this was possible, showed a loss of pressure in all instances. Replacement of the explanted AUS with a new device with the same size cuff and PRB in 14 of these 19 patients was successful in 12 (85.7%).ConclusionsThese results, and other theoretical considerations, suggest that recurrent incontinence, years after initially successful implantation of an AUS, is because of material failure of the PRB, probably attributable to its age and consequent loss of its ability to generate the pressure it was designed to produce, and that urethral atrophy does not occur. Simply replacing the old device with a new one with the same characteristics, unless there is a particular reason to do otherwise, is usually successful and avoids the complications of alternatives such as as cuff downsizing, implanting a PRB with a higher pressure range, implantation of a second cuff or transcorporeal cuff placement, all of which have been advocated in these patients.© 2015 The Authors BJU International © 2015 BJU International Published by John Wiley & Sons Ltd.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.