• Injury · Nov 2020

    Randomized Controlled Trial

    Stability of internal versus external fixation in osteoporotic pelvic fractures - a biomechanical analysis.

    • J Hack, Y Kranz, T Knauf, M Bäumlein, D Malcherczyk, S Ruchholtz, and L Oberkircher.
    • Center for Orthopaedics and Trauma Surgery, University Hospital Giessen and Marburg GmbH, Marburg, Germany. Electronic address: hackj@med.uni-marburg.de.
    • Injury. 2020 Nov 1; 51 (11): 2460-2464.

    IntroductionOsteoporosis-related fragility fractures of the pelvic ring (FFP) differ fundamentally from pelvic fractures in younger patients. However, very little is known about biomechanical stability of different osteosynthesis procedures addressing the anterior pelvic ring in these fractures. The aim of this study was to compare standard external fixation with internal fixation using a novel screw-and-rod system in osteoporotic fractures of the pelvic ring in terms of stiffness, plastic deformation and maximum load under cyclic loading in a human cadaveric model.Materials And MethodsA total of 18 embalmed osteoporotic cadaver pelvis specimens were randomized based on the T-score into a group for external fixation and a group for internal fixation. FFP type-IIB fractures were created. In addition to the external or internal fixator, a cement-augmented sacroiliac screw was implanted. Afterwards, axial cyclic loading was performed in a testing setup simulating one-leg stand.ResultsMean plastic deformation and stiffness both were significantly better in the internal fixation group than in the external fixation group (plastic deformation: 0.37 mm (SD: 0.23) versus 0.71 mm (SD: 0.26), p = 0.011; stiffness: 43.69 N/mm (SD: 18.39) versus 26.52 N/mm (SD: 9.76), p = 0.029). Maximum load did not differ significantly between internal fixator (506.3 N; SD: 129.4) and external fixator (461.1 N; SD: 147.4) (p = 0.515).ConclusionsSubmuscular internal fixation might be an interesting alternative to external fixation in clinical practice because of better biomechanical properties as well as several advantages in clinical use.Copyright © 2020. Published by Elsevier Ltd.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…