-
- Annemijn H Jonkman, Bhushan H Katira, Annia Schreiber, Cong Lu, Doreen Engelberts, Fernando Vieira, Alexandra Marquez, Arthur S Slutsky, Paul Dorian, and Laurent J Brochard.
- Keenan Research Centre, Li Ka Shing Knowledge Institute, St. Michael's Hospital, Toronto, Canada.
- Respir Care. 2021 Mar 1; 66 (3): 366377366-377.
BackgroundDuring the COVID-19 pandemic, a need for innovative, inexpensive, and simple ventilator devices for mass use has emerged. The Oxylator (CPR Medical Devices, Markham, Ontario, Canada) is an FDA-approved, fist-size, portable ventilation device developed for out-of-hospital emergency ventilation. It has not been tested in conditions of severe lung injury or with added PEEP. We aimed to assess the performance and reliability of the device in simulated and experimental conditions of severe lung injury, and to derive monitoring methods to allow the delivery of safe, individualized ventilation during situations of surge.MethodsWe bench-tested the functioning of the device with an added PEEP valve extensively, mimicking adult patients with various respiratory mechanics during controlled ventilation, spontaneous breathing, and prolonged unstable conditions where mechanics or breathing effort was changed at every breath. The device was further tested on a porcine model (4 animals) after inducing lung injury, and these results were compared with conventional ventilation modes.ResultsThe device was stable and predictable, delivering a constant flow (30 L/min) and cycling automatically at the inspiratory pressure set (minimum of 20 cm H2O) above auto-PEEP. Changes in respiratory mechanics manifested as changes in respiratory timing, allowing prediction of tidal volumes from breathing frequency. Simulating lung injury resulted in relatively low tidal volumes (330 mL with compliance of 20 mL/cm H2O). In the porcine model, arterial oxygenation, CO2, and pH were comparable to conventional modes of ventilation.ConclusionsThe Oxylator is a simple device that delivered stable ventilation with tidal volumes within a clinically acceptable range in bench and porcine lung models with low compliance. External monitoring of respiratory timing is advisable, allowing tidal volume estimation and recognition of changes in respiratory mechanics. The device can be an efficient, low-cost, and practical rescue solution for providing short-term ventilatory support as a temporary bridge, but it requires a caregiver at the bedside.Copyright © 2021 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.