Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single-Institution Case Series.
- Elmar M Delhaas, van AssemaDaniëlle M EDMEDepartment of Radiology and Nuclear Medicine, Erasmus University Medical Center, Rotterdam, The Netherlands., Alida C Fröberg, Ben G J C Zwezerijnen, Biswadjiet S Harhangi, FrankemaSander P GSPGCenter for Pain Medicine, Erasmus University Medical Center, Rotterdam, The Netherlands., HuygenFrank J P MFJPMCenter for Pain Medicine, Erasmus University Medical Center, Rotterdam, The Netherlands., and Aad van der Lugt.
- Center for Pain Medicine, Erasmus University Medical Center, Rotterdam, The Netherlands.
- Neuromodulation. 2021 Oct 1; 24 (7): 1190-1198.
BackgroundThe aim of this study was to assess the feasibility and diagnostic accuracy of an optimized 111 Indium-diethylenetriamine-penta-acetic-acid single-photon-emission computed tomography (CT) (111 In-DTPA SPECT-CT) examination in patients with suspected intrathecal drug delivery (ITDD) failure.Materials And MethodsRetrospective analysis of routinely collected observational data from a case series of patients in the setting of the academic Center for Pain Medicine, Departments of Radiology and Nuclear Medicine and Neurosurgery. Twenty-seven patients participated between January 2014 and January 2019. Thirty-six optimized examinations including standardized pump flow rate with additional SPECT-CT imaging and a stepwise standardized analysis were performed. A 10 mL mixture of medication and 20 MBq 111In-DTPA was injected into the pump reservoir. Planar and SPECT-CT images were acquired at 24, 48, and 72 hours (h) after injection and at 96 hours and/or seven days, if needed. All images were reassessed by the first two authors using an optimized procedure.Results And ConclusionsTwenty-two abnormalities were identified in 21 examinations, with these abnormalities consisting of leakage (n = 7), spinal catheter obstruction (n = 7), and cerebrospinal fluid flow obstruction (n = 8). Interventions (n = 19) confirmed the cause of ITDD failure. A false-positive finding at follow-up (n = 1) and a false-negative finding (n = 1) were encountered. Sensitivity was 95% (20/21) and the specificity 93% (14/15). A significant difference (p < 0.001) was found between the accuracy of the conventical and the optimized analysis. The optimized 111 In-DTPA SPECT-CT examination is a powerful diagnostic tool for detecting the cause of ITDD failure.© 2020 The Authors. Neuromodulation: Technology at the Neural Interface published by Wiley Periodicals LLC on behalf of International Neuromodulation Society.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.