• J Aerosol Med Pulm Drug Deliv · Oct 2011

    An in vitro assessment of aerosol delivery through patient breathing circuits used with medical air or a helium-oxygen mixture.

    • Andrew R Martin, Angela Ang, Ira M Katz, Sabine Häussermann, Georges Caillibotte, and Joëlle Texereau.
    • Medical Gases Group, Air Liquide Santé International, Centre de Recherche Claude-Delorme, Jouy-en-Josas, France. andrew-r.martin@airliquide.com
    • J Aerosol Med Pulm Drug Deliv. 2011 Oct 1; 24 (5): 225-34.

    BackgroundThe bench experiments presented herein were conducted in order to investigate the influence of carrier gas, either medical air or a helium-oxygen mixture (78% He, 22% O2), on the droplet size distribution and aerosol mass delivered from a vibrating mesh nebulizer through a patient breathing circuit.MethodsDroplet size distributions at the exit of the nebulizer T-piece and at the patient end of the breathing circuit were determined by laser diffraction. Additional experiments were performed to determine the effects on measured size distributions of gas humidity and of the droplet residence time during transport from the nebulizer to the laser diffraction measurement volume. Aerosol deposition in the nebulizer, breathing circuit, and on expiratory and patient filters was determined by photometry following nebulization of sodium fluoride solutions into the breathing circuit during simulated patient breathing.ResultsWith no humidification of the carrier gas, droplet volume median diameter (VMD) at the exit of the nebulizer T-piece was 5.5±0.1 μm for medical air, and 4.3±0.1 μm for helium-oxygen. Varying the aerosol residence time between the nebulizer and the measurement volume did not affect the measured size distributions; however, humidification of the carrier gases reduced differences in VMD at the nebulizer exit between medical air and helium-oxygen. At the patient end of the breathing circuit, droplet VMDs were 1.8±0.1 μm for medical air and 2.2±0.1 μm for helium-oxygen. The percentages of sodium fluoride recovered from the nebulizer, breathing circuit, patient filter, and expiratory filter were, respectively, 29.9±8.3, 40.4±5.6, 8.3±1.5, and 21.5±2.1% for air, and 32.6±2.2, 36.3±0.7, 12.0±1.4, and 19.1±1.1% for helium-oxygen.ConclusionsVentilation with helium-oxygen in place of air-oxygen mixtures can influence both the droplet size distribution and mass of nebulized aerosol delivered through patient breathing circuits. Assessment of these effects on aerosol delivery is important when incorporating helium-oxygen into patient ventilation strategies.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…