Paediatric anaesthesia · Dec 2020
Comparing oscillometric non-invasive and invasive intra-arterial blood pressure monitoring in term neonates under general anesthesia: A retrospective study.
- Tasuku Fujii and Kimitoshi Nishiwaki.
- Department of Anesthesiology, Nagoya University Hospital, Nagoya, Japan.
- Paediatr Anaesth. 2020 Dec 1; 30 (12): 1396-1401.
BackgroundOscillometric noninvasive blood pressure and/or invasive intra-arterial blood pressure are commonly used to measure the systolic, diastolic, and mean components of blood pressure. Agreement between the two methods has been reported in adults, children, and infants, but rarely in neonates, especially under general anesthesia.AimsThis retrospective study compared the agreement of each measured blood pressure value (oscillometric noninvasive or invasive intra-arterial blood pressure monitoring) in term neonates under general anesthesia.MethodsData were collected from neonates born at ≥36 weeks of gestation whose body weight was ≥2500 g and who underwent abdominal or noncardiac thoracic surgery with both oscillometric noninvasive and invasive intra-arterial blood pressure measurements from January 2015 to March 2020. The primary outcome was the agreement of systolic, diastolic, and mean blood pressure values between the two methods using Bland-Altman analysis.ResultsPaired blood pressure measurements (n = 1193) from 67 cases were compared. In Bland-Altman analysis, bias (standard deviation), 95% limits of agreement, and percentage error were -9.3 (8.4), -26.1-7.6, and 26.9% for systolic; 1.6 (6.5), -11.3-14.6, and 38.7% for diastolic; and -1.3 (5.8), -13.0-10.3, and 26.9% for mean blood pressure, respectively. During low blood pressure (intra-arterial mean blood pressure ≤30 mm Hg), the biases (standard deviation) of systolic, diastolic, and mean blood pressure were -11.4 (5.7), -0.7 (3.7), and -5.1 (4.2), whereas during high blood pressure (intra-arterial mean blood pressure ≥60 mm Hg), the values were 0.1 (9.7), 5.6 (9.4), and 6.4 (7.4), respectively.ConclusionsBased on the bias and percentage error, the mean blood pressure exhibited the most acceptable agreement between oscillometric noninvasive and invasive intra-arterial blood pressure monitoring in term neonates under general anesthesia. However, during hypertension or hypotension, there was a large discrepancy between the two methods.© 2020 John Wiley & Sons Ltd.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.