Case Reports
Inappropriate Defibrillator Shocks During Cervical Medial Branch Radiofrequency Ablation: A Case Report.
- Vasudha Goel, Hariharan Shankar, Siva K Mulpuru, and Harish Ramakrishna.
- From the Department of Anesthesiology, University of Minnesota, Minneapolis, Minnesota.
- A A Pract. 2020 Sep 1; 14 (11): e01286.
AbstractA 52-year-old woman with a history of cervical spondylosis, nonischemic dilated cardiomyopathy, and complete heart block with implantation of a cardiac resynchronization therapy defibrillator (Inogen XR, Boston Scientific, Natick, MA) underwent bilateral fluoroscopy-guided radiofrequency ablation of the medial branch nerves at C4, C5, and C6 levels at an ambulatory surgery center. After completion of the seemingly routine procedure, several alerts, including 1 inappropriate shock, were received via Boston Scientific's remote monitoring system. Tracings also showed that pacing inhibition occurred. When performing radiofrequency ablation in patients with a cardiac implantable electronic device, it is imperative to follow published best practice recommendations to minimize the risk of adverse events.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.