Comparative StudyA comparison of cardiac output by thoracic impedance and direct fick in children with congenital heart disease undergoing diagnostic cardiac catheterization.
- Katherine Taylor, Gustavo La Rotta, Brian W McCrindle, Cedric Manlhiot, Andrew Redington, and Helen Holtby.
- Department of Anesthesia, University of Toronto, Toronto, Ontario, Canada.
- J. Cardiothorac. Vasc. Anesth.. 2011 Oct 1;25(5):776-9.
Objective(S)To evaluate the measurement of cardiac output (CO) using continuous electrical bioimpedance cardiography (Physioflow; Neumedx, Philadelphia, PA) (CO(PF)) with a simultaneous direct Fick measurement (CO(FICK)) in children with congenital heart disease.DesignA prospective cohort study comparing 2 methods of measurement of CO.SettingA quaternary university-affiliated pediatric hospital.ParticipantsChildren undergoing cardiac catheterization for clinical care.InterventionsThe Physioflow measured continuous real time CO in 15-second epochs and simultaneous measurement of cardiac output by direct Fick (with mass spectrometry to assess VO(2)) were acquired.Measurements And Main ResultsSixty-five patients were recruited, and data from 56 (25 males) were adequate for analysis. The median age at study was 3.5 years (range, 0.4-16.6 years), and the median body surface area was 0.62 m(2) (range, 0.31-1.71). There were 25 of 56 (45%) with univentricular physiology. A total of 19,228 Physioflow data points were available for the analysis of which 14,569 (76%) were valid; 96% of the invalid measurements were identified as artifacts by the device. The average cardiac index of valid measurements was 3.09 ± 0.72 L/min/m(2). Compared with the Fick CO, the mean bias was -0.09 L/min, but the 95% limits of agreement were -3.20 to +3.01 L/min/m(2). Consequently, only 20 of 56 (36%) of measurements were within 20%, and 31 of 56 (55%) of measurements were within 30% of each other.ConclusionsCompared with measurements made by direct Fick, CO measured using the Physioflow device was unreliable in anesthetized children with congenital heart disease.Copyright © 2011 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.