• J Clin Monit Comput · Dec 2021

    Mitochondrial oxygen monitoring with COMET: verification of calibration in man and comparison with vascular occlusion tests in healthy volunteers.

    • R Ubbink, M A Wefers Bettink, W van Weteringen, and E G Mik.
    • Department of Anesthesiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands. r.ubbink@erasmusmc.nl.
    • J Clin Monit Comput. 2021 Dec 1; 35 (6): 1357-1366.

    AbstractMitochondria are the primary consumers of oxygen and therefore an important location for oxygen availability and consumption measurement. A technique has been developed for mitochondrial oxygen tension (mitoPO2) measurement, incorporated in the COMET. In contrast to most textbooks, relatively high average mitoPO2 values have been reported. The first aim of this study was to verify the validity of the COMET calibration for mitoPO2 measurements in human skin. The second aim was to compare the dynamics of mitoPO2 to several other techniques assessing tissue oxygenation. Firstly, we performed a two-point calibration. Mitochondrial oxygen depletion was achieved with vascular occlusion. A high mitoPO2 was reached by local application of cyanide. MitoPO2 was compared to the arterial oxygen partial pressure (PaO2). Secondly, for deoxygenation kinetics we compared COMET variables with the LEA O2C, SenTec OxiVenT™ and Medtronic INVOS™ parameters during a vascular occlusion test. 20 healthy volunteers were recruited and resulted in 18 datasets (2 times 9 subjects). The lowest measured mitoPO2 value per subject had a median [IQR] of 3.0 [1.0-4.0] mmHg, n = 9. After cyanide application the mitoPO2 was 94.1 mmHg [87.2-110.9] and did not differ significantly (n = 9, p = 0.5) from the PaO2 of 101.0 [98.0-106.0] mmHg. In contrast to O2C, OxiVenT™ and INVOS parameters, mitoPO2 declined within seconds with pressure on the probe. The kinetics from this decline are used to mitochondrial oxygen consumption (mitoVO2). This study validates the calibration of the COMET device in humans. For mitoVO2 measurements not only blood flow cessation but application of local pressure is of great importance to clear the measurement site of oxygen-carrying erythrocytes.© 2020. The Author(s).

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.