J Clin Monit Comput · Oct 2021
LetterEndotracheal tube inflation tubing defect: an unusual cause of intraoperative volume leak.
- Abhyuday Kumar, Amarjeet Kumar, Neeraj Kumar, and Ajeet Kumar.
- Department of Anesthesiology, AIIMS Patna, Room No. 15, OPD Block, Phulwarisarif, Patna, Bihar, India. drabhyu@gmail.com.
- J Clin Monit Comput. 2021 Oct 1; 35 (5): 1239-1240.
AbstractLoss of endotracheal cuff pressure can lead to airway compromise and loss of volume delivery in mechanical ventilation requiring immediate intervention. A 40 years old male posted for bilateral interpositional arthroplasty was intubated nasally with 7.5 mm flexometallic endotracheal tube. In the intraoperative period airway leak was detected due to loss of cuff pressure even after repeated attempt of cuff inflation. After changing endotracheal tube, leak was detected from the inflation tubing distal to the cuff, which was apparent only when cuff pressure was increased above 30 cm of H2O.© 2020. Springer Nature B.V.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.