• Spine · Jun 2000

    Clinical Trial

    Lumbar interbody fusion using the Brantigan I/F cage for posterior lumbar interbody fusion and the variable pedicle screw placement system: two-year results from a Food and Drug Administration investigational device exemption clinical trial.

    • J W Brantigan, A D Steffee, M L Lewis, L M Quinn, and J M Persenaire.
    • South Texas Orthopaedic and Spinal Surgery Associates, San Antonio, Texas 78240, USA.
    • Spine. 2000 Jun 1; 25 (11): 1437-46.

    Study DesignA carbon fiber-reinforced polymer cage implant filled with autologous bone was designed to separate the mechanical and biologic functions of posterior lumbar interbody fusion.ObjectivesTo test the safety and efficacy of the carbon cage with pedicle screw fixation in a 2-year prospective study performed at six centers under a protocol approved by the Food and Drug Administration, and to present the data supporting the Food and Drug Administration approved indications.Summary Of Background DataThe success of posterior lumbar interbody fusion has been limited by mechanical and biologic deficiencies of the donor bone. Some failures of pedicle screw fixation may be attributable to the absence of adequate load sharing through the anterior column. Combining an interbody fusion device with pedicle screw fixation may address some limitations of posterior lumbar interbody fusion or pedicle screw fixation in cases that are more complex mechanically.MethodsThis clinical study of posterior lumbar interbody fusion with pedicle screw fixation involved a prospective group of 221 patients.ResultsFusion success was achieved in 176 (98.9%) of 178 patients. In the management of degenerative disc disease in patients with prior failed discectomy surgery, clinical success was achieved in 79 (86%) of 92 patients, and radiographic bony arthrodesis in 91 (100%) of 91 patients. Disc space height, averaging 7.9 mm before surgery, was increased to 12.3 mm at surgery and maintained at 11.7 mm at 2 years. Fusion success was notdiminished over multiple fusion levels. These results were significantly better than those reported in prior literature. Although significant surgical complications occurred, those attributable to the implant devices occurred less frequently and generally were minor.ConclusionsThe Brantigan I/F Cage for posterior lumbar interbody fusion and the Variable Screw Placement System are safe and effective for the management of degenerative disc disease.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…