Randomized Controlled Trial Multicenter Study
Threshold-based insulin-pump interruption for reduction of hypoglycemia.
- Richard M Bergenstal, David C Klonoff, Satish K Garg, Bruce W Bode, Melissa Meredith, Robert H Slover, Andrew J Ahmann, John B Welsh, Scott W Lee, Francine R Kaufman, and ASPIRE In-Home Study Group.
- International Diabetes Center at Park Nicollet, Minneapolis, MN 55416, USA. richard.bergenstal@parknicollet.com
- N. Engl. J. Med.. 2013 Jul 18;369(3):224-32.
BackgroundThe threshold-suspend feature of sensor-augmented insulin pumps is designed to minimize the risk of hypoglycemia by interrupting insulin delivery at a preset sensor glucose value. We evaluated sensor-augmented insulin-pump therapy with and without the threshold-suspend feature in patients with nocturnal hypoglycemia.MethodsWe randomly assigned patients with type 1 diabetes and documented nocturnal hypoglycemia to receive sensor-augmented insulin-pump therapy with or without the threshold-suspend feature for 3 months. The primary safety outcome was the change in the glycated hemoglobin level. The primary efficacy outcome was the area under the curve (AUC) for nocturnal hypoglycemic events. Two-hour threshold-suspend events were analyzed with respect to subsequent sensor glucose values.ResultsA total of 247 patients were randomly assigned to receive sensor-augmented insulin-pump therapy with the threshold-suspend feature (threshold-suspend group, 121 patients) or standard sensor-augmented insulin-pump therapy (control group, 126 patients). The changes in glycated hemoglobin values were similar in the two groups. The mean AUC for nocturnal hypoglycemic events was 37.5% lower in the threshold-suspend group than in the control group (980 ± 1200 mg per deciliter [54.4 ± 66.6 mmol per liter] × minutes vs. 1568 ± 1995 mg per deciliter [87.0 ± 110.7 mmol per liter] × minutes, P<0.001). Nocturnal hypoglycemic events occurred 31.8% less frequently in the threshold-suspend group than in the control group (1.5 ± 1.0 vs. 2.2 ± 1.3 per patient-week, P<0.001). The percentages of nocturnal sensor glucose values of less than 50 mg per deciliter (2.8 mmol per liter), 50 to less than 60 mg per deciliter (3.3 mmol per liter), and 60 to less than 70 mg per deciliter (3.9 mmol per liter) were significantly reduced in the threshold-suspend group (P<0.001 for each range). After 1438 instances at night in which the pump was stopped for 2 hours, the mean sensor glucose value was 92.6 ± 40.7 mg per deciliter (5.1 ± 2.3 mmol per liter). Four patients (all in the control group) had a severe hypoglycemic event; no patients had diabetic ketoacidosis.ConclusionsThis study showed that over a 3-month period the use of sensor-augmented insulin-pump therapy with the threshold-suspend feature reduced nocturnal hypoglycemia, without increasing glycated hemoglobin values. (Funded by Medtronic MiniMed; ASPIRE ClinicalTrials.gov number, NCT01497938.).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.