• Int. J. Pediatr. Otorhinolaryngol. · Jan 2017

    The utility of a handheld metal detector in detection and localization of pediatric metallic foreign body ingestion.

    • Javan Nation and Wen Jiang.
    • Rady Children's Hospital San Diego Pediatric Otolaryngology, 3030 Children's Way Suite 402, San Diego, CA 92123, USA; University of California San Diego, Division of Otolaryngology, 200 West Arbor Drive, San Diego, CA 92103, USA. Electronic address: jnation@ucsd.edu.
    • Int. J. Pediatr. Otorhinolaryngol. 2017 Jan 1; 92: 1-6.

    ObjectiveTo test the ability of a handheld metal detector (HHMD) to identify the presence and location of ingested metallic foreign bodies (MFBs) in children.MethodsProspective case series enrolling children suspected of metallic foreign body ingestion presenting to the Emergency Department. Thirty-eight children were enrolled and the HHMD was used to detect the presence and location of a MFB. Results were compared to standard radiographic studies.ResultsThirty-seven of the 38 ingested foreign bodies were MFBs. Of the 37 MFBs, the HHMD positively identified 33, and 4 were missed by HHMD but identified on radiography. When positive, the location indicated by HHMD correlated 100% with radiograph. There were 33 true positives, 0 false positives, 4 false negatives, and 1 true negative. This resulted in a sensitivity of 89% (95% CI of 75%-96%) and specificity of 100% (95% CI of 2.5%-100%).ConclusionOur study demonstrates the accuracy of HHMD in the identification and localization of metallic foreign bodies. We propose an emergency room foreign body protocol that uses HHMD as an early screening tool in triage in order to expedite the process of obtaining Otolaryngology consultation and potentially shorten the wait time to the operating room or discharge. In instances were outside films are previously performed, HHMD use may be able to minimize the overall radiation exposure to children by obviating the need for repeat radiographs. As the sensitivity is not 100%, a negative HHMD screening does not negate the need for a standard radiograph in order to avoid missed MFBs. HHMD is best suited for detection of coins, which accounts for the majority of the MFB ingestions, and may not be suitable for all metallic objects since the amount of metal may decrease its sensitivity.Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…