    Evaluation of centrifugal blood pumps in term of hemodynamic performance using simulated neonatal and pediatric ECMO circuits.

    • Shigang Wang, Morgan K Moroi, Allen R Kunselman, John L Myers, and Akif Ündar.
    • Department of Pediatrics, Penn State Health Pediatric Cardiovascular Research Center, Penn State Milton S. Hershey Medical Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
    • Artif Organs. 2020 Jan 1; 44 (1): 16-27.

    AbstractThe objective of this translational study was to evaluate the FDA-approved PediMag, CentriMag, and RotaFlow centrifugal blood pumps in terms of hemodynamic performance using simulated neonatal and pediatric extracorporeal membrane oxygenation (ECMO) circuits with different sizes of arterial and venous cannulae. Cost of disposable pump heads was another important variable for this particular study. The experimental circuit was composed of one of the centrifugal pump heads, a polymethylpentene membrane oxygenator, neonatal and pediatric arterial/venous cannulae, and 1/4-inch ID tubing. Circuits were primed with lactated Ringer's solution and packed human red blood cells (hematocrit 35%). Trials were conducted at 36°C using the three pump heads and different cannulae (arterial/venous cannulae: 8 Fr/18 Fr, 10 Fr/20 Fr, and 12 Fr/22 Fr) at various flow rates (200-2400 mL/min, 200 mL/min increments) and rotational speeds. Pseudo patient pressure was 60 mm Hg. Real-time pressure and flow data were recorded for analysis. The RotaFlow pump had a higher pressure head and flow range compared with the PediMag and CentriMag pumps at the same rotational speed and identical experimental settings (P < 0.001). The PediMag pump had lower flow output than others (P < 0.001). Small-caliber arterial cannulae and higher flow rates predictably created higher circuit pressures and pressure drops. There was no significant difference in hemodynamic energy delivered to the pseudo patient with each of the three pumps. The arterial cannula had the highest pressure drop and hemodynamic energy loss in the circuit when compared to the oxygenator and arterial tubing. The RotaFlow centrifugal pump had a significantly better hemodynamic performance when compared to the PediMag and CentriMag blood pumps at identical experimental conditions in simulated neonatal and pediatric ECMO settings. In addition, the cost of the RotaFlow pump head ($400) is 20 to 30-fold less than the other centrifugal pumps [CentriMag ($12 000) or PediMag ($8000)] that were evaluated in this translational study.© 2019 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.