    Biomechanical testing of a PEEK-based dynamic instrumentation device in a lumbar spine model.

    • Christian Herren, Agnes Beckmann, Sabine Meyer, Miguel Pishnamaz, Marion Mundt, Rolf Sobottke, Andreas Prescher, Marcus Stoffel, Bernd Markert, Philipp Kobbe, Hans-Christoph Pape, Peer Eysel, and Jan Siewe.
    • Department for Trauma and Reconstructive Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany. Electronic address: cherren@ukaachen.de.
    • Clin Biomech (Bristol, Avon). 2017 May 1; 44: 67-74.

    BackgroundThe purpose of this study was to investigate the range-of-motion after posterior polyetheretherketone-based rod stabilisation combined with a dynamic silicone hinge in order to compare it with titanium rigid stabilisation.MethodsFive human cadaveric lumbar spines with four vertebra each (L2 to L5) were tested in a temperature adjustable spine-testing set-up in four trials: (1) native measurement; (2) kinematics after rigid monosegmental titanium rod instrumentation with anterior intervertebral bracing of the segment L4/5; (3) kinematics after hybrid posterior polyetheretherketone rod instrumentation combined with a silicone hinge within the adjacent level (L3/4) and (4) kinematics after additional decompression with laminectomy of L4 and bilateral resection of the inferior articular processes (L3). During all steps, the specimens were loaded quasi-statically with 1°/s with pure moment up to 7.5Nm in flexion/extension, lateral bending and axial rotation.FindingsIn comparison to the native cadaveric spine, both the titanium device and polyetheretherketone-based device reduce the range-of-motion within the level L4/5 significantly (flexion/extension: reduction of 77%, p<0.001; lateral bending: reduction of 62%, p<0.001; axial rotation: reduction of 71%, p<0.001). There was a clear stabilisation effect after hybrid-instrumentation within the level L3/4, especially in flexion/extension (64%, p<0.001) and lateral bending (62%, p<0.001) but without any effect on the axial rotation. Any temperature dependency has not been observed.InterpretationSurprisingly, the hybrid device compensates for laminectomy L4 and destabilising procedure within the level L3/4 in comparison to other implants. Further studies must be performed to show its effectiveness regarding the adjacent segment instability.Copyright © 2017 Elsevier Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…