Comparative Study
Reliability of Intraoperative Knee Range of Motion Measurements by Goniometer Compared with Robot-Assisted Arthroplasty.
- Hyuck Min Kwon, Ick-Hwan Yang, Woo-Suk Lee, Alvin Ray L Yu, Sang Yun Oh, and Kwan Kyu Park.
- Department of Orthopedic Surgery, Yonsei University College of Medicine, Seoul, Korea.
- J Knee Surg. 2019 Mar 1; 32 (3): 233-238.
AbstractAccurate measurement of knee range of motion (ROM) is critical to predict the outcomes of knee surgery and prognosis. We investigated the reliability of knee ROM measurements by goniometer compared with robotic system. Fifty-three patients with medial osteoarthritis who were planning to undergo unicompartmental knee arthroplasty (UKA) with robotic UKA were prospectively enrolled. During the operation, knee ROM measurement was performed in both flexion and extension before and after insertion of the implant using both a goniometer and robotic system. The intraclass correlation coefficient (ICC) of extension measured by the goniometer and robotic system showed good agreement; however, the ICC of flexion did not show good agreement. During passive flexion, the mean values measured before insertion of the implant were significantly lower by goniometer (134.6 ± 6.43) than by robot (145.4 ± 6.80; p = 0.017); likewise, the mean values after insertion of the implant were significantly lower by goniometer (138.6 ± 6.07) than by robotic system (147.0 ± 6.60; p = 0.045). A goniometer can underestimate knee ROM measurements compared with robotic system, especially in flexion. Orthopaedic surgeons should be cautious when measuring the flexion angle with a goniometer.Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.