• Heart Rhythm · May 2014

    Randomized Controlled Trial Multicenter Study

    A pacemaker transthoracic impedance sensor with an advanced algorithm to identify severe sleep apnea: the DREAM European study.

    • Pascal Defaye, Ines de la Cruz, Julio Martí-Almor, Roger Villuendas, Paul Bru, Jérémie Sénéchal, Renaud Tamisier, and Jean-Louis Pépin.
    • Arrhythmia Unit, Cardiology Department, University Hospital, Grenoble, France. Electronic address: PDefaye@chu-grenoble.fr.
    • Heart Rhythm. 2014 May 1; 11 (5): 842-8.

    BackgroundSleep apnea (SA) is associated with cardiovascular diseases and is highly prevalent in patients with pacemakers (PMs).ObjectiveTo validate a transthoracic impedance sensor with an advanced algorithm (sleep apnea monitoring) for identifying severe SA.MethodsPatients with indications for PM (VVI/DDD) were enrolled regardless of symptoms suggesting SA. Severe SA diagnosis was acknowledged when the full polysomnography gave an apnea-hypopnea index (PSG-AHI) of ≥30 events/h. The PSG-AHI was compared with the respiratory disturbance index evaluated by the SAM algorithm (SAM-RDI) compiled from the device during the same diagnosis night, and the performance of the device and the SAM algorithm was calculated to identify patients with severe SA. The agreement between methods was assessed by using Bland and Altman statistics.ResultsForty patients (mean age 73.8 ± 19.1 years; 67.5% men; body mass index 27.7 ± 4.4 kg/m(2)) were included. Severe SA was diagnosed by PSG in 56% of the patients. We did not retrieve SAM-RDI data in 14% of the patients. An optimal cutoff value for the SAM-RDI at 20 events/h was obtained by a receiver operator characteristic curve analysis, which yielded a sensitivity of 88.9% (95% confidence interval [CI] 65.3%-98.6%), a positive predictive value of 88.9% (95% CI 65.3%-98.6%), and a specificity of 84.6% (95% CI 54.6%-98.1%) (n = 31). The Bland-Altman limits of agreement for PSG-AHI (in events per hour) were [-14.1 to 32.4].ConclusionThe results suggest that an advanced algorithm using PM transthoracic impedance could be used to identify SA in patients with PMs outside the clinic or at home.Copyright © 2014 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…