Randomized Controlled Trial Multicenter Study Comparative Study
Assessment of a scalp EEG-based automated seizure detection system.
- K M Kelly, D S Shiau, R T Kern, J H Chien, M C K Yang, K A Yandora, J P Valeriano, J J Halford, and J C Sackellares.
- Center for Neuroscience Research, Allegheny-Singer Research Institute, Allegheny General Hospital, Pittsburgh, PA 15212, USA.
- Clin Neurophysiol. 2010 Nov 1; 121 (11): 1832-43.
ObjectiveThe purpose of this study was to evaluate and validate an offline, automated scalp EEG-based seizure detection system and to compare its performance to commercially available seizure detection software.MethodsThe test seizure detection system, IdentEvent™, was developed to enhance the efficiency of post-hoc long-term EEG review in epilepsy monitoring units. It translates multi-channel scalp EEG signals into multiple EEG descriptors and recognizes ictal EEG patterns. Detection criteria and thresholds were optimized in 47 long-term scalp EEG recordings selected for training (47 subjects, ∼3653h with 141 seizures). The detection performance of IdentEvent was evaluated using a separate test dataset consisting of 436 EEG segments obtained from 55 subjects (∼1200h with 146 seizures). Each of the test EEG segments was reviewed by three independent epileptologists and the presence or absence of seizures in each epoch was determined by majority rule. Seizure detection sensitivity and false detection rate were calculated for IdentEvent as well as for the comparable detection software (Persyst's Reveal®, version 2008.03.13, with three parameter settings). Bootstrap re-sampling was applied to establish the 95% confidence intervals of the estimates and for the performance comparison between two detection algorithms.ResultsThe overall detection sensitivity of IdentEvent was 79.5% with a false detection rate (FDR) of 2 per 24h, whereas the comparison system had 80.8%, 76%, and 74% sensitivity using its three detection thresholds (perception score) with FDRs of 13, 8, and 6 per 24h, respectively. Bootstrap 95% confidence intervals of the performance difference revealed that the two detection systems had comparable detection sensitivity, but IdentEvent generated a significantly (p<0.05) smaller FDR.ConclusionsThe study validates the performance of the IdentEvent™ seizure detection system.SignificanceWith comparable detection sensitivity, an improved false detection rate makes the automated seizure detection software more useful in clinical practice.Copyright © 2010 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.