Critical care medicine · Jul 1996
An intrinsic positive end-expiratory pressure lung model, with and without flow limitation.
- K L Yang and C Wang.
- Division of Pulmonary and Critical Care Medicine, University of Louisville, KY 40202, USA.
- Crit. Care Med. 1996 Jul 1; 24 (7): 1261-5.
ObjectiveTo design an intrinsic positive end-expiratory pressure (PEEP) lung model that has the property of air flow limitation.DesignMechanical lung model study of intrinsic PEEP.SettingLung models were set up in the research laboratory.InterventionsIntrinsic PEEP lung models were created with and without flow limitation. In the model with flow limitation, intrinsic PEEP was created by replacing a portion of the expiratory circuit with a collapsible Penrose tube and by placing this portion the circuit under water. The expiratory circuit became a part of the respiratory airway, with flow limitation occurring at the Penrose drain. In the model without flow limitation, intrinsic PEEP was generated with a fixed linear resistor, which was inserted in the expiratory circuit to produce a similar level of intrinsic PEEP. Multiple levels of external PEEP, both above and below the initial intrinsic PEEP, were applied.Measurements And Main ResultsAt each level of external PEEP, peak airway pressure, plateau airway pressure, isovolume air flow, internal lung pressure, and intrinsic PEEP were measured. Peak airway pressure, plateau pressure, and internal lung pressure were minimally affected if the external PEEP was less than the intrinsic PEEP in the lung model with flow limitation. Intrinsic PEEP was reduced with external PEEP. However, if intrinsic PEEP was induced without dynamic airway closure or flow limitation, any level of external PEEP caused an immediate increase in peak airway pressure, plateau airway pressure, and internal lung pressure and a decrease in isovolume flow. External PEEP has little effect on the levels of intrinsic PEEP.ConclusionsWe demonstrated two different models of an intrinsic PEEP lung model. The interactions between intrinsic PEEP and externally applied PEEP were different. The lung model with collapsible tube closely simulated the human respiratory system with flow limitation. This lung model may be useful for the future study of intrinsic PEEP and pulmonary mechanics.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.