- H Y Arandia and P H Byles.
- Can Anaesth Soc J. 1981 Sep 1; 28 (5): 467-70.
AbstractThe feasibility of applying positive end expiratory pressure within the Bain circuit using an Emerson PEEP valve was studied. It was determined that varying degrees of PEEP may be applied within the Bain circuit using an Emerson PEEP valve during either mechanically or manually controlled ventilation but not during spontaneous breathing. There was some loss of measured tidal volume with increasing PEEP, due primarily to compression gas losses but also from leakage throughout the system. The maximum gas loss was 100 ml per minute at PEEP 15 cm H2O under test conditions. PEEP may be applied between the Bain tubing and the manifold. However, one should then never allow the patient to breathe spontaneously as the PEEP valve presents an obstruction to inspiratory flow. With the PEEP valve relocated between the Bain manifold and the ventilator hose it was found that application of PEEP was possible without this disadvantage, but only during mechanical ventilation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.