• Spine · Jul 2015

    Case Reports

    Analysis of the Fracture Mechanism of Ti-6Al-4V Alloy Rods that failed Clinically after Spinal Instrumentation Surgery.

    • Kenta Yamanaka, Manami Mori, Ken Yamazaki, Ruriko Kumagai, Minoru Doita, and Akihiko Chiba.
    • *Institute for Materials Research, Tohoku University, Sendai, Japan †Department of Materials and Environmental Engineering, Sendai National College of Technology, Natori, Japan; and ‡Department of Orthopedic Surgery, School of Medicine, Iwate Medical University, Morioka, Japan.
    • Spine. 2015 Jul 1; 40 (13): E767-73.

    Study DesignRetrieval analysis of 2 Ti-6Al-4V alloy rods that fractured after spinal instrumentation surgery.ObjectiveTo determine the mechanism that underlies fractures of Ti-6Al-4V alloy rods after spinal instrumentation surgery from a materials science viewpoint.Summary Of Background DataRod failures after spinal instrumentation surgery are often reported and many case-based studies have been published. However, the details of the mechanism that underlies the fractures have not yet been fully elucidated.MethodsTwo patients, a 71-year-old female and an 11-year-old male, underwent radiography and removal of their fractured rods. The latter patient had been treated using the growing-rod method. Metallurgical failure analysis of the retrieved rods was conducted, and material properties were compared between the unused and fractured rods.ResultsThe microstructures and mechanical properties of the Ti-6Al-4V alloy rods that failed after spinal instrumentation surgery were similar to those of unused rods. Analysis of the fracture surfaces clearly identified fatigue cracking in both cases that would have lowered the resistance of the rods to failures caused by external stresses. Shot blasting the surfaces of Ti-6Al-4V alloy rods and bending the rods to fit particular contours, which is always conducted during spinal instrumentation surgery, probably introduced fatigue cracking because the alloy is highly notch sensitive.ConclusionImprovements should be made to rod design and/or rod material, because the fatigue resistance of titanium alloys is intrinsically lower than that of other commercially available rod materials, including cobalt-chromium alloys. These imperfections may have greater consequences for the growing-rod method and pseudarthrosis, where the rods are not completely fixed, and they subsequently suffer from severe long-arm moments.Level Of EvidenceN/A.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…