- Egidijus Kontautas, Andrej Pijadin, Andrius Vilkauskas, and Aurelijus Domeika.
- Department of Orthopedics and Traumatology, Medical Academy, Lithuanian University of Health Sciences, Kaunas, Lithuania. egidijuskon@yahoo.com
- Medicina (Kaunas). 2012 Jan 1; 48 (2): 80-3.
UnlabelledThe aim of this study was to evaluate and compare the biomechanical effects of locking reconstruction plate positioning on the osteosynthesis of clavicle midshaft simulated transverse fractures.Material And MethodsTwelve synthetic clavicles with simulated midshaft transverse fractures were repaired with a 3.5-mm locking reconstruction plate in the anteroinferior or the superior position. The clavicles were randomly assigned to 2 groups (6 per group). Each repaired clavicle was tested in cantilever bending by using the universal testing machine. The maximal load and the displacement of the specimens at a load of 40 N were recorded for each group.ResultsThe anteroinferior plating osteosynthesis with a 3.5-mm locking reconstruction plate could bear an average maximal load of 183.3 N (SD, 11.3); the corresponding load for the superior plating osteosynthesis with the identical implants was 444.8 N (SD, 102.3), and the mean displacement was 1.5 mm (SD, 0.5) and 0.7 mm (SD, 0.2), respectively.ConclusionsThe superior plating osteosynthesis of simulated midshaft transverse clavicle fractures with the 7-hole 3.5-mm locking reconstruction plate had a significantly higher bending (from top to bottom) load to failure in comparison with the anteroinferior plating osteosynthesis of the clavicle with the identical implants. Clavicles plated with the 7-hole 3.5-mm locking reconstruction plate at the superior aspect exhibited a significantly greater biomechanical stability at a load of 40 N than those plated at the anteroinferior aspect.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.