Comparative StudyMonitoring intraoperative neuromuscular blockade and blood pressure with one device (TOF-Cuff): A comparative study with mechanomyography and invasive blood pressure.
- G Veiga Ruiz, J García Cayuela, J Orozco Montes, M Parreño Caparrós, B García Rojo, and J L Aguayo Albasini.
- Servicio de Anestesiología, Reanimación y T. del dolor, Hospital General Universitario Santa Lucía, Cartagena, Murcia, España. Electronic address: gloriaveiga@gmail.com.
- Rev Esp Anestesiol Reanim. 2017 Dec 1; 64 (10): 560-567.
ObjectiveThe overall objective of the study is to determine the ability of TOF-Cuff device (blood-pressure modified cuff, including stimulation electrodes) to monitor with the same device the non-invasive blood pressure (NIBP) and the depth of a neuromuscular blockade (NMB) induced pharmacologically, by stimulation of the brachial plexus at the humeral level and recording evoked changes in arterial pressure.Material And MethodClinical, single-centre, open-controlled study with 32 adult patients ASA I-III for scheduled elective surgery under general anaesthesia in supine position, for the validation of neuromuscular monitoring, comparing the values obtained from neuromuscular relaxation TOF-Cuff with those obtained by mechanomyography (MMG) (control method) during the recovery phase of NMB, when a TOF ratio>0.7 and>0.9 (primary endpoint) were reached respectively. And an additional consecutive study of 17 patients for validation of NIBP monitoring with TOF-Cuff device vs invasive blood pressure measured by an intra-arterial catheter. All data were analyzed using the Bland-Altman method.ResultsRecovery from NMB measured with the TOF-Cuff was earlier compared to MMG. Comparing TOF-ratio>0.9 measured with TOF-Cuff vs TOF-ratio>0.7 with MMG, a specificity of 91% and a positive predictive value of 84% were obtained. In NIBP measurement, the mean error and standard deviation of both systolic blood pressure (1.6±7mmHg) and diastolic blood pressure (-3.4±6.3) were within the European accuracy requirements for medical devices.ConclusionsThe TOF-Cuff device has been shown to be valid and safe in the monitoring of NMB and in the measurement of NIBP, with no patient presenting any adverse events, skin-level lesions or residual pain. It is not interchangeable with MMG, having a TOF-ratio>0.9 quantified by the TOF-Cuff device, a good correlation with a TOF-ratio>0.7 on MMG.Copyright © 2017 Sociedad Española de Anestesiología, Reanimación y Terapéutica del Dolor. Publicado por Elsevier España, S.L.U. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.