J Spinal Disord Tech · Apr 2010
Pedicle screw augmentation with polyethylene tape: a biomechanical study in the osteoporotic thoracolumbar spine.
- Takahiko Hamasaki, Nobuhiro Tanaka, JinHwan Kim, Motohiro Okada, Mitsuo Ochi, and William C Hutton.
- Department of Orthopaedic Surgery, Emory Spine Center, Atlanta, GA 30329, USA.
- J Spinal Disord Tech. 2010 Apr 1; 23 (2): 127-32.
Study DesignBiomechanical study using human thoracolumbar spines.ObjectiveTo assess the strength of fixation of pedicle screws augmented with Nesplon tape applied sublaminar or applied subpars.Summary Of Background DataThere are no reports on the biomechanical assessment of Nesplon tape used for pedicle screw augmentation.MethodsExperiment 1: pedicle screws were inserted into bilateral pedicles as follows: (1) pedicle screw alone connected to a rod (PS alone), (2) pedicle screw augmented with Nesplon tape applied sublaminar, connected to a rod [pedicle screw applied sublaminar (PSSL)], and (3) pedicle screw augmented with tape applied subpars, connected to a rod [pedicle screw applied subpars (PSSP)]. The rods were pulled and pushed until the pedicle screw was loose. Experiment 2: 6 thoracolumbar spines were biomechanically tested as follows: axial compression, flexion, extension, left and right lateral bending, and left and right axial rotation. This sequence was applied to: (1) the intact spine; (2) the spine made unstable; (3) the spine fixed by pedicle screws and rods (PS alone); (4) the same as 3, but with the pedicle screws augmented using Nesplon tape applied sublaminar (PSSL); and (5) the same as 3, but with pedicle screws augmented using tape applied subpars (PSSP). From the load-deformation curves, stiffness values were calculated.ResultsExperiment 1: the pedicle screws augmented by sublaminar or subpars tape (PSSL or PSSP) showed significantly greater fixation strength compared with those pedicles that were not augmented by tape (PS alone). Experiment 2: a pedicle screw/rod construct augmented by tape provided a stiffer construct than the same construct without augmentation.ConclusionsPedicle screws that are inserted into the osteoporotic thoracolumbar spine and augmented by Nesplon tape applied sublaminar or subpars provide firmer fixation of the screws and a stiffer pedicle screw/rod construct than the same construct without augmentation by tape.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.