• J Occup Environ Hyg · Jan 2016

    Performance of an improperly sized and stretched-out loose-fitting powered air-purifying respirator: Manikin-based study.

    • Shuang Gao, Roy T McKay, Michael Yermakov, Jinyong Kim, Tiina Reponen, Xinjian He, Kazushi Kimura, and Sergey A Grinshpun.
    • a Center for Health-Related Aerosol Studies, Department of Environmental Health, University of Cincinnati , Cincinnati , Ohio.
    • J Occup Environ Hyg. 2016 Jan 1; 13 (3): 169-76.

    AbstractThe objective of this study was to investigate the protection level offered by a Powered Air-Purifying Respirator (PAPR) equipped with an improperly sized or stretched-out loose-fitting facepiece using constant and cyclic flow conditions. Improperly sized PAPR facepieces of two models as well as a stretched-out facepiece were tested. These facepieces were examined in two versions: with and without exhaust holes. Loose-fitting facepieces (size "large") were donned on a small manikin headform and challenged with sodium chloride (NaCl) aerosol particles in an exposure chamber. Four cyclic flows with mean inspiratory flows (MIFs) of 30, 55, 85, and 135 L/min were applied using an electromechanical Breathing Recording and Simulation System (BRSS). The manikin Fit Factor (mFF) was determined as the ratio of aerosol concentrations outside (Cout) to inside (Cin) of the facepiece, measured with a P-Trak condensation particle counter (CPC). Results showed that the mFF decreased exponentially with increasing MIF. The mFF values of the stretched-out facepiece were significantly lower than those obtained for the undamaged ones. Facepiece type and MIF were found to significantly affect the performance of the loose-fitting PAPR. The effect of the exhaust holes was less pronounced and depended on the facepiece type. It was concluded that an improperly sized facepiece might potentially offer relatively low protection (mFF < 250) at high to strenuous workloads. The testing was also performed at a constant inhalation flow to explore the mechanism of the particle-facepiece interaction. Results obtained with cyclic flow pattern were consistent with the data generated when testing the loose-fitting PAPR under constant flow conditions. The time-weighted average values of mFF calculated from the measurements conducted under the constant flow regime were capable of predicting the protection under cyclic flow regime. The findings suggest that program administrators need to equip employees with properly sized facepieces and remove stretched-out ones from workplace. Manufacturers should emphasize the importance of proper sizing with their user instructions.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.