• J Clin Sleep Med · Mar 2016

    Pressure-Relief Features of Fixed and Autotitrating Continuous Positive Airway Pressure May Impair Their Efficacy: Evaluation with a Respiratory Bench Model.

    • Kaixian Zhu, Sami Aouf, Gabriel Roisman, and Pierre Escourrou.
    • Air Liquide Healthcare, Gentilly, France.
    • J Clin Sleep Med. 2016 Mar 1; 12 (3): 385-92.

    Study ObjectivesPressure-relief features are aimed at improving the patient's comfort during continuous positive airway pressure (CPAP) treatment for obstructive sleep apnea. The objective of this study was to determine the effect of these therapy features on fixed CPAP and autotitrating CPAP (APAP) treatment efficacy.MethodsSeven pressure-relief features applied by three CPAP devices were included in our study (Remstar Auto: C-Flex 3, C-Flex+ 3, A-Flex 3, P-Flex; AirSense 10: EPR 3; Prisma 20A: SoftPAP 2 and 3). In fixed CPAP, the devices were subjected to a 10-min bench-simulated obstructive apnea sequence (initial apnea-hypopnea index, AHI = 60/h) with and without pressure-relief features. In APAP, the sequence was lengthened to 4.2 h (initial AHI = 58.6/h). The residual AHI and mean/median pressure were compared with and without pressure-relief features.ResultsCompared to conventional CPAP, where pressure was adjusted to be just sufficient to control the simulated obstructive events, C-Flex+ 3, P-Flex, and EPR 3 failed to normalize the breathing flow and did not reduce the AHI. The mean pressures with the three features, respectively, were 1.8, 2.6, and 2.6 cmH2O lower than the conventional CPAP. Compared to conventional APAP, similar levels of control were observed with pressure-relief features, apart from P-Flex where the delivered mean pressure was lower and residual AHI greater. The device-reported mean/median pressures in APAP with A-Flex 3, P-Flex, EPR 3, and SoftPAP 3 were higher than that measured on the bench.ConclusionsPressure-relief features may attenuate CPAP efficacy if not adjusted for at the time of their introduction. In clinical practice, efficacy can be ensured by increasing the therapeutic pressure delivered by fixed CPAP or by enabling the pressure-relief features prior to initial pressure titration. Device-reported pressures in APAP devices with pressure relief activated may overstate delivered pressures.© 2016 American Academy of Sleep Medicine.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.