    Comparison of Accu Chek Inform II point-of-care test blood glucose meter with Hexokinase Plasma method for a diabetes mellitus population during surgery under general anesthesia.

    • B Lascaris, H G Freling, M A Edens, M J Fokkert, C G Olthof, and R J Slingerland.
    • Department of Anesthesiology, University Medical Center Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands. b.lascaris@umcg.nl.
    • J Clin Monit Comput. 2022 Apr 1; 36 (2): 355-361.

    PurposeBlood glucose (BG) concentrations of patients with diabetes mellitus (DM) are monitored during surgery to prevent hypo- and hyperglycemia. Access to point-of-care test (POCT) glucose meters at an operating room will usually provide monitoring at shorter intervals and may improve glycemic control. However, these meters are not validated for patients under general anesthesia.MethodsThis cross-sectional study included 75 arterial BG measurements from 75 patients (71 with DM, mostly insulin dependent) who underwent elective non-cardiac surgery under general anesthesia. Arterial blood samples were taken at least 60 minutes after induction. One drop of blood was used for Accu Chek Inform II (ACI II) POCT BG meter and the residual blood was sent to the clinical laboratory for a Hexokinase Plasma reference method. A Bland-Altman plot was used to visualize the differences between both methods, and correlation was assessed using the intra-class correlation coefficient (ICC).ResultsThe results showed an estimated mean difference of 0.8 mmol/L between ACI II and the reference method, with limits of agreement equal to -0.6 and 2.2 mmol/L. In general, the reference method produced higher values than ACI II. ICC was 0.955 (95% CI 0.634-0.986), P < 0.001, and concordance correlation coefficient (CCC) was 0.955 (95% CI 0.933-0.970).ConclusionArterial BG measurements during surgery in patients with DM under general anesthesia using POCT BG meter are in general lower than laboratory measurements, but the ICC and CCC show a clinically acceptable correlation. We conclude that POCT measurements conducted on arterial specimens using the ACI II provide sufficiently accurate results for glucose measurement during surgery under general anesthesia.© 2021. The Author(s).

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…