Comparative StudyElectrostatic discharges and their effect on the validity of registered values in intracranial pressure monitors.
- Morten Andresen, Marianne Juhler, and Ole Cornelius Thomsen.
- Clinic of Neurosurgery, Copenhagen University Hospital, Copenhagen; and.
- J. Neurosurg.. 2013 Nov 1;119(5):1119-24.
ObjectIntracranial pressure (ICP) monitoring is used extensively in clinical practice, and as such, the accuracy of registered ICP values is paramount. Clinical observations of nonphysiological changes in ICP have called into question the accuracy of registered ICP values. Subsequently, the authors have tried to determine if the ICP monitors from major manufacturers were affected by electrostatic discharges (ESDs), if the changes were permanent or transient in nature, and if the changes were modified by the addition of different electrical appliances normally used in the neurointensive care unit environment.MethodsThe authors established a test setup in the neurointensive care unit using a large container filled with isotonic saline, creating a phantom patient. Intracranial pressure monitors were sequentially lowered into the container and subjected to a predefined test battery of ESDs.ResultsFive pressure monitors from 4 manufacturers were evaluated. Three monitors containing electrical circuitry at the tip of the transducer were all affected by ESDs. Clinically significant permanent changes in the reported ICP values for 1 pressure monitor were observed, as well as temporary deflections for 2 other monitors. The monitors had different levels of sensitivity to discharges at low voltages.ConclusionsThese results explain some of the sudden shifts in ICP noted in the clinical setting. However, a clear deflection pattern related to the addition of electrical appliances was not found. The authors recommend instituting policies for reducing the risk of subjecting patients to ESDs in the neurointensive care unit setting.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.