• Clin. Orthop. Relat. Res. · Sep 2015

    Are Complications Associated With the Repiphysis(®) Expandable Distal Femoral Prosthesis Acceptable for Its Continued Use?

    • Eric L Staals, Marco Colangeli, Nikolin Ali, José M Casanova, Davide M Donati, and Marco Manfrini.
    • Department of Orthopaedics - Musculoskeletal Oncology, Istituto Ortopedico Rizzoli (IOR) Clinica 3, Via Pupilli 1, 40136, Bologna, Italy, eric.staals@ior.it.
    • Clin. Orthop. Relat. Res. 2015 Sep 1; 473 (9): 3003-13.

    BackgroundReconstruction of the distal femur after resection for malignant bone tumors in skeletally immature children is challenging. The use of megaprostheses has become increasingly popular in this patient group since the introduction of custom-made, expandable devices that do not require surgery for lengthening, such as the Repiphysis(®) Limb Salvage System. Early reports on the device were positive but more recently, a high complication rate and associated bone loss have been reported.Questions/PurposesWe asked: (1) what are the clinical outcomes using the Musculoskeletal Tumor Society (MSTS) scoring system after 5-year minimum followup in patients treated with this prosthesis at one center; (2) what are the problems and complications associated with the lengthening procedures of this implant; and (3) what are the specific concerns associated with revision of this implant?MethodsAt our institute, between 2002 and 2007, the Repiphysis(®) expandable prosthesis was implanted in 15 children (mean age, 8 years; range, 6-11 years) after distal femoral resection for malignant bone tumors. During this time, the general indication for use of this implant was resection of the distal femur for localized malignant bone tumors in pediatric patients. Alternative techniques used for this indication were modular prosthetic reconstruction, massive (osteoarticular or intercalary) allograft reconstruction, or rotationplasty. Age and tumor extension were the main factors to decide on the surgical indication. Of the 15 patients who had this prosthesis implanted during reconstruction surgery, five died with the implant in situ or underwent amputation before 5 years followup and the remaining 10 were evaluated at a minimum of 5 years (mean, 104 months; range, 78-140 months). No patients were lost to followup. These 10 patients were long-term survivors and underwent the lengthening program. They were included in our study analysis. The first seven lengthening procedures were attempted in an outpatient setting; however, owing to pain and burning sensations experienced by the patients, the procedures failed to achieve the desired lengthening. Therefore, other procedures were performed with the patients under general anesthesia. We reviewed clinical data at index surgery for all 15 patients. We further analyzed the lengthening procedures, implant survival, radiographic and functional results, for the 10 long-term survivors. Functional results were assessed according to the MSTS scoring system. Complications were classified according to the International Society of Limb Salvage (ISOLS) classification system.ResultsNine of the 10 survivors underwent revision of the implant for mechanical failure. They had a mean MSTS score of 64% (range, 47%-87%) before revision surgery. At final followup the 10 long-term surviving patients had an average MSTS score of 81% (range, 53%-97%). In total, we obtained an average lengthening of 39 mm per patient (range, 17-67 mm). Exact expansion of the implant was unpredictable and difficult to control. Nine of 10 of the long-term surviving patients underwent revision surgery of the prosthesis-eight for implant breakage and one for stem loosening. At revision surgery, six patients had another type of expandable prosthesis implanted and three had an adult-type megaprosthesis implanted. In five cases, segmental bone grafts were used during revision surgery to compensate for loss of bone stock.ConclusionsWe could not comfortably expand the Repiphysis(®) prosthesis in an outpatient setting because of pain experienced by the patients during the lengthening procedures. Furthermore, use of the prosthesis was associated with frequent failures related to implant breakage and stem loosening. Revisions of these procedures were complex and difficult. We no longer use this prosthesis and caution others against the use of this particular prosthesis design.Level Of EvidenceLevel IV, therapeutic study.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…