-
Invest. Ophthalmol. Vis. Sci. · Apr 2013
Comparative Study Clinical TrialComparative analysis of repeatability of manual and automated choroidal thickness measurements in nonneovascular age-related macular degeneration.
- Sieun Lee, Nader Fallah, Farzin Forooghian, Ashley Ko, Kaivon Pakzad-Vaezi, Andrew B Merkur, Andrew W Kirker, David A Albiani, Mei Young, Marinko V Sarunic, and Mirza Faisal Beg.
- School of Engineering Science, Simon Fraser University, Burnaby, British Columbia, Canada. leeau@sfu.ca
- Invest. Ophthalmol. Vis. Sci. 2013 Apr 23; 54 (4): 2864-71.
PurposeWe compared the reproducibility and mutual agreement of the subfoveal choroidal thickness measurements by expert raters and an automated algorithm in enhanced depth imaging optical coherence tomography (EDI-OCT) images of eyes with nonneovascular age-related macular degeneration (AMD).MethodsWe recruited 44 patients with nonneovascular AMD and EDI-OCT images were acquired. Subfoveal choroidal thickness was measured manually by two expert raters and automatically by a graph-cut-based algorithm. Drusen area was measured using the automated software (version 6) of Cirrus SD-OCT. The manual and automated choroidal thickness measurements were compared in reproducibility, mutual agreement, and correlation with drusen area.ResultsThe mean subfoveal choroidal thickness was 246 ± 63 μm for the first rater, 214 ± 68 for the second rater, and 209 ± 53 for the automated algorithm. Intraclass correlation coefficients (ICC) and 95% confidence intervals (CI) were 0.96 (CI 0.94-0.98) between the raters, 0.85 (CI 0.77-0.90) between the first rater and the automated algorithm, and 0.84 (CI 0.75-0.89) between the second rater and the automated algorithm. Repeat scan measurement ICCs were 0.91 (CI 0.86-0.94) for the first rater, 0.96 (CI 0.94-0.97) for the second rater, and 0.87 (CI 0.80-0.92) for the automated algorithm. Manual and automated measurements were correlated with drusen area.ConclusionsThe automated algorithm generally yielded smaller choroidal thickness than the raters with a moderate level of agreement. However, its repeat scan measurement repeatability was comparable to that of the manual measurements. The mean difference between the raters indicated possible biases in different raters and rating sessions. The correlation of the automated measurements with the drusen area was comparable to that of the manual measurements. Automated subfoveal choroidal thickness measurement has potential use in clinical practice and clinical trials, with possibility for reduced time and labor cost.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.